All measurements are subject to error. To some
extent, measuring instruments have built-in or inherent errors, called systematic errors. (For example, a kitchen scale might
consistently yield results that are 25 g too high or a thermometer a reading
that is 2°C too low.) Limitations in an experimenter’s skill or ability to read
a scientific instrument also lead to errors and give results that may be either
too high or too low. Such errors are called random errors.
Precision refers to the degree of reproducibility
of a measured quantity that is, the closeness of agreement when the same
quantity is measured several times. The precision of a series of measurements
is high (or good) if each of a series of measurements deviates by only a small
amount from the average. Conversely, if there is wide deviation among the
measurements, the precision is poor (or low). Accuracy refers to how close a measured value is to the
accepted, or actual value. High-precision measurements are not always accurate,
a large systematic error could be present. (A tight cluster of three darts near
the edge of a dart board can be considered precise but not very accurate if the
intention was to strike the center of the board.) Still, scientists generally
strive for high precision in measurements.
To illustrate these ideas, consider measuring the
mass of an object by using the two balances shown in the figure. One of the
balances is a single- pan balance that gives the mass in grams with only one
decimal place. The other balance is a sophisticated analytical balance that
gives the mass in grams with four decimal places. The accompanying table gives
results obtained when the object is weighed three times on each balance. For
the single-pan balance, the average of the measurements is 10.5 g, with
measurements ranging from 10.4 g to 10.6 g. For the analytical balance, the
average of the measurements is 10.4978 g, with measurements ranging from
10.4977 g to 10.4979 g. The scatter in the data obtained with the single- pan
balance (±0.1 g) is greater than that obtained with the analytical balance (±
0.0001 g). Thus, the results obtained by using the single-pan balance have
lower (or poorer) precision than those obtained by using the analytical
balance.
No comments:
Post a Comment